Product data sheet
Product Characteristics
Total number of inputs and outputs
Number of digital outputs: 2; Number of analog outputs: 1
Measuring range
0.2.... .50 l/min
0.01.... .3 m³/h .3 m³/h
System connections
Threaded connection G 3/4 DN20 Flat seal
APPLICATIONS
Special properties
Gold-plated contacts
Applications
Adder function; for industrial applications
Mounting
Pipe connection via adapter
Medium
conductive liquids; water; water as basic medium
Description of the medium
Conductivity: ≥ 20 µS/cm
Viscosity: < 70 mm²/s (40 °C)
Temperature of the medium [°C]
-10.... .70
Compressive strength [bar]
16
Compressive strength [MPa]
1.6
Maximum permissible working pressure (for applications complying with CRN standards) [bar]
11.2
Electrical data
Operating voltage [V]
18.... .30 DC; (according to SELV/PELV)
Current loss [mA]
95; (24 V)
Protection class
III
Inverted phase protection
yes
Power-on delay time [s]
5
Total inputs/outputs
Total number of inputs and outputs
Number of digital outputs: 2; Number of analog outputs: 1
Inputs
Inputs
Counter reset
Outputs
Number of outputs
2
Output signals
Switch signal; Analog signal; Pulse signal; IO-Link; (Configurable)
Electrical design
PNP/NPN
Number of digital outputs
2
Output Function
Normally open/closed; (parameterizable)
Switching output DC voltage drop maximum value [V]
2
Continuous current load of switching output DC [mA]
200
Number of analog outputs
1
Analog current output [mA]
4.... .20; (adjustable range)
Maximum load value [Ω]
500
Analog voltage output [V]
0.... .10; (adjustable range)
Minimum load resistance [Ω]
2000
Pulse output
Flow meter
Short-circuit protection
yes
Short-circuit protection type
Pulse
Overload protection
Yes
Measuring/setting range
Measuring range
0.2.... .50 l/min
0.01.... .3 m³/h
Display range
-60... .60 l/min .60 l/min
-3.6... .3.6 m³/h .3.6 m³/h
Resolution
0.1 l/min
0.001 m³/h
Switching point, SP
0.5..... .50 l/min
0.027....3 m³/h .3 m³/h
Recovery point, rP
0.2... .49.8 l/min .49.8 l/min
0.012... .2.985 m³/h .2.985 m³/h
Measured value starting point, ASP
0.... .40 l/min
0.... .2.4 m³/h
Measured value end point, AEP
10.... .50 l/min
0.6... .3 m³/h .3 m³/h
Steps
0.1 l/min
0.001 m³/h
Flow monitoring
Pulse value
0.00001.... .50 000 m³
Pulse length [s]
0,005.... .2
Temperature monitoring
Measuring range [°C]
-20.... .80
Resolution [°C]
0.2
Switching point, SP [°C]
-19.2.... .80
Recovery point, rP [°C]
-19.6.... .79.6
Measured value starting point [°C]
-20.... .60
Measured value end point [°C]
0.... .80
Setting step [°C]
0.2
Accuracy / Deviation
Flow monitoring
Accuracy (in the measuring range)
± (0,8 % MW + 0,5 % MEW)
Repeatability
± 0,2 % MEW
Temperature monitoring
Accuracy [K]
± 2,5 (Q > 5 l/min)
Reaction time
Flow rate monitoring
Reaction time [s]
0.15; (dAP = 0, T19)
Programmable delay time dS, dr [s]
0.... .50
Damping process value dAP [s]
0.... .5
Temperature monitoring
Response time T05 / T09 [s]
T09 = 20 (Q > 5 l/min)
Software / Programming
Parameterization
Flow monitoring; Quantity meter; Preset capacity meter; Temperature monitoring; Hysteresis/window; Normally open/closed; Output polarity; Current/voltage/pulse output; Start-up delay; Display can be switched off; Display units
Interfaces
Communication interfaces
IO-Link