Product data sheet
Control Function – AX460 Only
Controller Type
P, PI, PID (Configurable)
Control Outputs
Analog
Current output control (0 to 100%)
Time proportioning cycle time
5 to 60s, programmable in 1s steps displayed with doP or
roP
Pulse frequency
1 to 120 pulses per minute
Controller action
Reverse, direct or bi-directional (programmable)
Proportional band
0.1 to 1000%, programmable in 0.1% increments
Integral action time (Reset)
1 to 7200s in 1s steps, 0 = Off
Derivative
0.1 to 1000s in 0.1s increments
Auto/Manual
User-programmable
Tags: